The GW Instek GPM-8213 Power Meter provides as many as 19 power measurement parameters, including voltage (Vrms/ V+pk / V-pk), current (Irms/ I+pk / I-pk), frequency (VHz/ IHz), power (P/ P+pk / P-pk), crest factor (CFV/ CFI), apparent power (VA), reactive power (VAR), power factor (PF), phase angle (DEG), total harmonic distortion (THDV/ THDI), high-accuracy voltage/current/power measurement capabilities (reading: ±0.1%; level: ±0.1%). The advantages of TFT LCD have been efficiently deployed to simple mode and standard mode. Simple mode displays conventional power meter’s four measurement parameters to meet the requirement of accuracy and clarity for the test on manufacturing process. Standard mode extends the display to the maximum of 8 measurement parameters (2 major measurements + 6 monitor measurements) to satisfy the various measurement application requirements of R&D, design, and quality verification.
Ideal for Standby Power Consumption Testing
For device Under Test (DUT) requiring IEC 62301/EN 50564 standby power consumption test, GPM-8213 provides the optimal measurement supports, including test frequency bandwidth of DC~6kHz, the minimum current level of 5mA (resolution: 0.1uA), power measurement resolutions (1uW for minimum current and voltage levels; 1mW for maximum current and voltage levels), crest factor reaching 3 (half range reaching 6), and measurement of total harmonic distortion (at least 13th order power harmonic). For large voltage /large current measurement applications of general power measurement, GPM-8213 provides PT / CT rate function to collocate with external potential transformer or current transformer to meet the measurement requirements.
With respect to data retrieval and storage, the standard RS-232C/ USB interfaces (virtual COM)/ LAN can be utilized to edit and retrieve programs or the optional GPIB interface (installed by manufacturer) can be selected to meet users’ automatic test system requirements.
GW Instek GPM-8213 Power Meter Features:
-
4” TFT LCD
-
Two data display modes
-
Standard : two major measurement items + six secondary measurement items
-
Simple : displaying test data of four different measurement items
-
Met the requirement of IEC 62301 power measurement
-
Voltage/ current test frequency bandwidth : DC~6kHz
-
Watt resolution : 1mW
-
Lowest current range : 5mA, resolution : 0.1uA
-
Current/voltage measurements reach CF=3 for distorted wave and CF=6 for half range
-
W-h (power vs time)/A-h (current vs time) integration function
-
Total harmonic distortion measurement
-
Front panel test terminal
-
Standard interface : RS-232C, USB device, LAN
-
Optional interface : GPIB
-
Optional test fixture : GPM-001 (Available with Universal or Europe socket)
GW Instek GPM-8213 Power Meter Specifications:
|
●The specifications apply when the GPM is warmed up for at least 30 minutes. ●Below are the basic conditions required to operate the GPM within specifications:
|
Specification Condition |
Temperature: 23˚C±5˚C Humidity: <80%RH(non-condensing) |
Operation Condition |
Temperature0°C~ 40°C,
|
Storage Condition |
Temperature −40°C ~70°C Humidity < 90%RH(non-condensing) |
Power Source |
AC 100-240V, 50–60Hz ; Consumption Max. 25VA |
Dimension & Weight |
270(W) x 110(H) x 350(D) mm, Approx. 2.9kg |
Input
Item |
Spec. |
||
Input voltage |
600 Vrms |
||
Input current |
20 Arms |
||
Input impedance(50/60 Hz) |
Voltage |
2.4MΩ |
|
Current | 5mA – 200mA |
500mΩ |
|
0.5A – 20A |
5mΩ |
||
Maximum display voltage |
700 Vrms |
||
Maximum display current |
25 Arms |
||
Maximum allowable isolation voltage |
300 V |
||
Low frequency filter |
Cut-off frequency |
500 Hz |
Display
Synchronization frequency | 45Hz~ 6kHz |
Average | 1, 2, 4, 8, 16, 32, 64 |
Displayed items(Standard mode) | 8 items simultaneously. |
Displayed items(Simple mode) | 4 items simultaneously. |
Displayed digits | 5 |
Voltage converter (PT ratio) | 1 to 9999.999 |
Current converter (CT ratio) | 1 to 9999.999 |
Measurement items | Voltage, current, active power, apparent power, reactive power, power factor, phase angle, frequency, integrated current, integrated power, positive integrated power, negative integrated power, integration time, voltage crest factor, current crest factor, voltage peak, current peak, Thd |
Displayed measurement parameters | Vdc, Vrms, V+pk, V-pk, Idc, Irms, I+pk, I-pk, P, P+pk, P-pk, VA, VAR, PF, CFV, CFI, DEG, VHz, IHz, THDV, THDI |
Voltage Measurement
Measurement range | CF=3 : | 15V, 30V, 60V, 150V, 300V, 600V | |
CF=6 : | 7.5V, 15V, 30V, 75V, 150V, 300V | ||
Crest factor | 3, 6 | ||
Accuracy | Effective range | 1 % to 105 % of range | |
DC | ±(0.2 % reading + 0.2 % range) | ||
45 Hz ≤ f ≤ 66 Hz | ±(0.1 % reading + 0.1 % range) | ||
66 Hz < f ≤ 1kHz | ±(0.1 % reading + 0.2 % range) | ||
1 kHz < f ≤ 6kHz | ± 3 % of range | ||
The filter is turned on | Increase 0.3 % reading@ 45Hz to 66Hz | ||
Temperature effect | 5-18˚C / 28-40˚C | Increase ±0.03% reading /˚C | |
Residual noise | 0.5 % of range |
Current Measurement
Measurement range | CF=3 : | 5mA, 10mA, 20mA, 50mA, 100mA, 200mA, 500mA, 1A, 2A, 5A, 10A, 20A | |
CF=6 : | 2.5mA, 5mA, 10mA, 25mA, 50mA, 100mA, 250mA, 0.5A, 1A, 2.5A, 5A, 10A | ||
Crest factor | 3, 6 | ||
Accuracy | Effective range | 1 % to 105 % of range | |
DC | ±(0.2 % reading + 0.2 % range) | ||
45 Hz ≤ f ≤ 66 Hz | ±(0.1 % reading + 0.1 % range) | ||
66 Hz < f ≤ 1kHz | ±(0.1 % reading + 0.2 % range) | ||
1 kHz < f ≤ 6kHz | ± 3 % of range | ||
The filter is turned on | Increase 0.3 % reading@ 45Hz to 66Hz | ||
Temperature effect | 5-18˚C / 28-40˚C | Increase ±0.03% reading /˚C | |
Residual noise | 0.5 % of range |
Power Measurement
Accuracy | Effective range | 1 % to 110 % of range |
DC | ±(0.2 % reading + 0.2 % range) | |
45 Hz ≤ f ≤ 66 Hz | ±(0.1 % reading + 0.1 % range) | |
66 Hz < f ≤ 1kHz | ±(0.1 % reading + 0.3 % range) | |
1 kHz < f ≤ 6kHz | ± 3 % of range | |
The filter is turned on | Increase 0.3 % reading@ 45Hz to 66Hz | |
Temperature effect | 5-18˚C / 28-40˚C | Increase ±0.03% reading /˚C |
Frequency Measurement
Measurement range | The filter is turned on | 30.000Hz to 499.99Hz |
The filter is turned off | 30.000Hz to 9.9999kHz | |
Measurement items | Voltage, Current | |
Effective input range | 10% to 105% of voltage input range | |
Accuracy | ±(0.06 % reading) |
Integrator Measurement
Integrator | Accuracy | ±(Accuracy of voltage or current+ 0.1 % reading) |
Time | Range | 0 hour 0 minute to 9999 hours 59 minutes |
Accuracy | ±0.01% ±1second |