These units beat every other comparable dry-well in the industry in performance, size, weight, convenience, ease of calibration, software, and price.
In addition, the heating and cooling rate of each of these dry-wells is adjustable from the front panel, thermal switches can be checked for actuation testing, and multiple-hole inserts are available for a variety of probe sizes.
Fluke dry-wells are easy to calibrate. You don’t even have to open the case. This means less maintenance costs and less down time when they do need calibration.
The Interface-it software lets you adjust set-points and ramp rates, log dry-well readings to a file, create an electronic strip chart, and perform thermal switch testing with data collection. The software is written for Windows and has a great graphical interface. Regardless of whether you want basic software or a completely automated calibration system, Fluke have what you want.
Every dry-well shipped is tested at the factory, and every unit comes with a NIST-traceable calibration. There’s no extra charge for the report, because we consider it an essential ingredient in our quality program. You shouldn’t have to pay extra for calibration procedures we perform anyway.
Fluke 9103
The 9103 covers below-ambient temperatures as low as –25 °C. The 9103 is stable to ±0.02 °C, and its display is calibrated to an accuracy of ±0.25 °C at all temperatures within its range. In just eight minutes, 0 °C is reached, and 100 °C is reached in six minutes, so your time is spent calibrating—not waiting.
The 9103 reaches temperatures 50 °C below ambient, so –25 °C is reached under normal ambient conditions. Our competitors like to advertise their units as reaching –45 °C when they really mean –45 °C below ambient, which typically means it will go to –20 °C. Our unit does not require you to work in a walk-in freezer to achieve its full advertised range.
Choose one of three removable inserts sized for probes from 1/16 inch to 1/2 inch in diameter. Insert A handles a full range of probe sizes with a single well of each size. Insert B features two wells each of 3/8, 1/4, and 3/16 inches in diameter for doing comparison calibrations. Insert C has six 1/4-inch-diameter wells for multiple probe calibrations, and Insert D has three pairs of metric sized wells.
Fluke 9140
The 9140 has a temperature range of 35 °C to 350 °C, and it reaches its maximum temperature in 12 minutes. At six pounds, it’s small enough to easily carry in one hand. It’s truly a unique innovation in dry-wells.
The unit has a stability of ±0.05 °C or better and a uniformity of at least 0.4 °C in the largest-diameter wells and 0.1 °C in the smaller wells. Despite its small size, this unit performs.
Use the display, calibrated to ±0.5 °C, as your reference, or use an external thermometer for maximum calibration accuracy. With three removable inserts to choose from, the 9140 is as versatile as it is fast.
Fluke 9103 User Manual