PRESYS MCS-XV Universal Multifunction Process Calibrator

Presys MCS-XV Universal Multifunction Process Calibrator Features:

  • Measures and generates mA, mV, Volts, ohms, RTD, TC, and Hz with accuracy 0.01% up to full scale.
  • Up to four optional modules of 250 mmH2O pressure to 10,000 psi. It has the barometric reference (optional) for absolute pressure indication.
  • Hart Full Configurator (optional). Compatible with DD configuration files recorded in FieldComm Group.
  • 24Vdc power supply for two-wire transmitter. internal 250 ohms resistor configurable.
  • Documentation capability. Automatic calibration tasks can be created or executed with or without the use of computers.
  • Calibration Reports issued directly to USB printer or PDF file.
  • Data logger function for data acquisition and display in graphical format.
  • Touchscreen display with high visibility friendly interface.
  • Compact, portable, powered by rechargeable batteries for field use.
  • Also available in versions Desktop (DT) and 19″ Rack Mounting (RM).
  • e-Connected: Ethernet, Wi-Fi (optional), Pen-drive, Hart, Profibus*, Fieldbus Foundation*, Modbus* and USB Host/Device

*Check availability

    Presys MCS-XV Datasheet

    Presys MCS-XV User Manual

 

If you would like more information regarding this product, simply email us here.
  • This field is for validation purposes and should be left unchanged.

The PRESYS MCS-XV Process Multifunction Calibrator enables measurement and generation of signals used in instrumentation and process control. It provides all the features required for easier calibration and adjustment services on process instruments. It has high levels of accuracy, including aspects relating to changes in room temperature and maintains the specifications over long periods of time.

The PRESYS MCS-XV Process Multifunction Calibrator is designed for field use and includes: carrying case, holders with fastener rings or belts for a hands-free operation, 5.7” display with LED backlight for a better viewing in poor lighting conditions, rechargeable battery and large memory capacity to store the values obtained during calibrations enabling the transfer of these to the PC when needed.

In addition to these features the PRESYS MCS-XV is available in a 19″ Rack Mount version (RM) or as a Desktop Instrument (DT).

The PRESYS MCS-XV Universal Multifunction Process Calibrator incorporates state-of-the-art concepts of automatic calibration and adjustment via computer. Instrument and computer share measurement data obtained for a more efficient handling of information such as report and certificate issues, automatic work management, data sorting and storage, for an overall coverage of quality procedure requirements, especially those related to ISO-9000.

The PRESYS MCS-XV has HART® communication for reading and setting parameters of field devices that have this protocol and can be supplied with up to four pressure sensors. Thus, a single calibrator can have different ranges of pressure, for example vacuum, 0 to 100 psi, 0 to 1,000 psi and 0 to 3,000 psi, or any combination of available ranges.

Presys MCS-XV Portable version in use.

Presys MCS-XV Universal Multifunction Process Calibrator Specifications:

Ranges Resolution Accuracy Remarks
Specifications – Inputs
milivolt -150 to 150 mV 0.001 mV ± 0.01 % FS *** Rinput > 10 MΩ
-500 to -150 mV 0.01 mV ± 0.02 % FS auto-ranging
150 to 2450 mV 0.01 mV ± 0.02 % FS
volt -10 to 45 V 0.0001 V ± 0.02 % FS Rinput > 1 MΩ
mA -5 to 24.5 mA 0.0001 mA ± 0.02 % FS Rinput < 120Ω
resistance 0 to 400 Ω 0.01 Ω ± 0.01 % FS Excitation current 0.85 mA
400 to 2500 Ω 0.01 Ω ± 0.03 % FS auto-ranging
frequency* 0 to 600 Hz 0.01 Hz ± 0.04 Hz Rinput > 50KΩ
600 to 1300 Hz 0.1 Hz ± 0.2 Hz Voltage DCmax = 30 V
1300 to 5000 Hz 1 Hz ± 2 Hz AC Signal from 0.3 to 30 V
auto-ranging
counter* 0 to 108 -1 count 1 count The same remark as frequency Pulses Frequency < 3000 Hz
Pt-100 -200 to 850 °C / -328 to 1562 °F 0.01 °C / 0.01 °F ± 0.1 °C / ± 0.2 °F IEC-60751
Pt-1000 -200 to 400 °C / -328 to 752 °F 0.1 °C / 0.1 °F ± 0.1 °C / ± 0.2 °F IEC-60751
Cu-10 -200 to 260 °C / -328 to 500 °F 0.1 °C / 0.1 °F ± 2.0 °C / ± 4.0 °F Minco 16-9
Ni-100 -60 to 250 °C / -76 to 482 °F 0.1 °C / 0.1 °F ± 0.2 °C / ± 0.4 °F DIN-43760
probe** -200 to 850 °C / -328 to 1562 °F 0.01 °C / 0.01 °F ± 0.1 °C / ± 0.2 °F IEC-60751
TC-J -210 to 1200 °C / -346 to 2192 °F 0.1 °C / 0.1 °F ± 0.2 °C / ± 0.4 °F IEC-60584
TC-K -270 to -150 °C / -454 to -238 °F 0.1 °C / 0.1 °F ± 0.5 °C / ± 1.0 °F IEC-60584
-150 to 1370 °C / -238 to 2498 °F 0.1 °C / 0.1 °F ± 0.2 °C / ± 0.4 °F
TC-T -260 to -200 °C / -436 to -328 °F 0.1 °C / 0.1 °F ± 0.6 °C / ± 1.2 °F IEC-60584
-200 to -75 °C / -328 to -103 °F 0.1 °C / 0.1 °F ± 0.4 °C / ± 0.8 °F
-75 to 400 °C / -103 to 752 °F 0.1 °C / 0.1 °F ± 0.2 °C / ± 0.4 °F
TC-B 50 to 250 °C / 122 to 482 °F 0.1 °C / 0.1 °F ± 2.5 °C / ± 5.0 °F IEC-60584
250 to 500 °C / 482 to 932 °F 0.1 °C / 0.1 °F ± 1.5 °C / ± 3.0 °F
500 to 1200 °C / 932 to 2192 °F 0.1 °C / 0.1 °F ± 1.0 °C / ± 2.0 °F
1200 to 1820 °C / 2192 to 3308 °F 0.1 °C / 0.1 °F ± 0.7 °C / ± 1.4 °F
TC-R -50 to 300 °C / -58 to 572 °F 0.1 °C / 0.1 °F ± 1.0 °C / ± 2.0 °F IEC-60584
300 to 1760 °C / 572 to 3200 °F 0.1 °C / 0.1 °F ± 0.7 °C / ± 1.4 °F
TC-S -50 to 300 °C / -58 to 572 °F 0.1 °C / 0.1 °F ± 1.0 °C / ± 2.0 °F IEC-60584
300 to 1760 °C / 572 to 3200 °F 0.1 °C / 0.1 °F ± 0.7 °C / ± 1.4 °F
TC-E -270 to -150 °C / -454 to -238 °F 0.1 °C / 0.1 °F ± 0.3 °C / ± 0.6 °F IEC-60584
-150 to 1000 °C / -238 to 1832 °F 0.1 °C / 0.1 °F ± 0.1 °C / ± 0.2 °F
TC-N -260 to -200 °C / -436 to -328 °F 0.1 °C / 0.1 °F ± 1.0 °C / ± 2.0 °F IEC-60584
-200 to -20 °C / -328 to -4 °F 0.1 °C / 0.1 °F ± 0.4 °C / ± 0.8 °F
-20 to 1300 °C / -4 to 2372 °F 0.1 °C / 0.1 °F ± 0.2 °C / ± 0.4 °F
TC-L -200 to 900 °C / -328 to 1652 °F 0.1 °C / 0.1 °F ± 0.2 °C / ± 0.4 °F DIN-43710
TC-C 0 to 1500 °C / 32 to 2732 °F 0.1 °C / 0.1 °F ± 0.5 °C / ± 1.0 °F W5Re / W26Re
1500 to 2320 °C / 2732 to 4208 °F 0.1 °C / 0.1 °F ± 0.7 °C / ± 1.4 °F
(*) Accuracy valid since the frequency output is not configured.
(**) The Probe is a separate input used as reference thermometer. The related accuracy is relative only to the MCS-XV.
(***) FS = Full Scale.
Specifications – Outputs
milivolt -10 to 110 mV 0.001 mV ± 0.02 % FS *** Rout < 0.3 Ω
volt -0.5 to 12 V 0.0001 V ± 0.02 % FS Rout < 0.3 Ω
mA 0 to 24 mA 0.0001 mA ± 0.02 % FS Rmax = 700 Ω
2-wire transmitter (XTR) 4 to 24 mA 0.0001 mA ± 0.02 % FS Vmax = 60 V
resistance 0 to 400 Ω 0.01 Ω ± 0.02 % FS For external excitation current
of 1.0 mA
0 to 2500 Ω 0.1 Ω ± 0.03 % FS
frequency 0 to 100 Hz 0.01 Hz ± 0.02 Hz Peak value:
0 to 10000 Hz 1 Hz ± 2 Hz 22 V / 25 mA max.
pulse 0 to 108 – 1 pulse 1 pulse Peak value:
22 V / 25 mA max.
Pulses frequency up to 10000 Hz
Pt-100 -200 to 850 °C / -328 to 1562 °F 0.01 °C / 0.01 °F ± 0.2 °C / ± 0.4 °F IEC-60751
Pt-1000 -200 to 400 °C / -328 to 752 °F 0.1 °C / 0.1 °F ± 0.1 °C / ± 0.2 °F IEC-60751
Cu-10 -200 to 260 °C / -328 to 500 °F 0.1 °C / 0.1 °F ± 2.0 °C / ± 4.0 °F Minco 16-9
Ni-100 -60 to 250 °C / -76 to 482 °F 0.1 °C / 0.1 °F ± 0.2 °C / ± 0.4 °F DIN-43760
TC-J -210 to 1200 °C / -346 to 2192 °F 0.1 °C / 0.1 °F ± 0.4 °C / ± 0.8 °F IEC-60584
TC-K -270 to -150 °C / -454 to -238 °F 0.1 °C / 0.1 °F ± 1.0 °C / ± 2.0 °F IEC-60584
-150 to 1370 °C / -238 to 2498 °F 0.1 °C / 0.1 °F ± 0.4 °C / ± 0.8 °F
TC-T -260 to -200 °C / -436 to -328 °F 0.1 °C / 0.1 °F ± 1.2 °C / ± 2.4 °F IEC-60584
-200 to -75 °C / -328 to -103 °F 0.1 °C / 0.1 °F ± 0.8 °C / ± 1.6 °F
-75 to 400 °C / -103 to 752 °F 0.1 °C / 0.1 °F ± 0.4 °C / ± 0.8 °F
TC-B 50 to 250 °C / 122 to 482 °F 0.1 °C / 0.1 °F ± 5.0 °C / ± 10.0 °F IEC-60584
250 to 500 °C / 482 to 932 °F 0.1 °C / 0.1 °F ± 3.0 °C / ± 6.0 °F
500 to 1200 °C / 932 to 2192 °F 0.1 °C / 0.1 °F ± 2.0 °C / ± 4.0 °F
1200 to 1820 °C / 2192 to 3308 °F 0.1 °C / 0.1 °F ± 1.4 °C / ± 2.8 °F
TC-R -50 to 300 °C / -58 to 572 °F 0.1 °C / 0.1 °F ± 2.0 °C / ± 4.0 °F IEC-60584
300 to 1760 °C / 572 to 3200 °F 0.1 °C / 0.1 °F ± 1.4 °C / ± 2.8 °F
TC-S -50 to 300 °C / -58 to 572 °F 0.1 °C / 0.1 °F ± 2.0 °C / ± 4.0 °F IEC-60584
300 to 1760 °C / 572 to 3200 °F 0.1 °C / 0.1 °F ± 1.4 °C / ± 2.8 °F
TC-E -270 to -150 °C / -454 to -238 °F 0.1 °C / 0.1 °F ± 0.6 °C / ± 1.2 °F IEC-60584
-150 to 1000 °C / -238 to 1832 °F 0.1 °C / 0.1 °F ± 0.2 °C / ± 0.4 °F
TC-N -260 to -200 °C / -436 to -328 °F 0.1 °C / 0.1 °F ± 2.0 °C / ± 4.0 °F IEC-60584
-200 to -20 °C / -328 to -4 °F 0.1 °C / 0.1 °F ± 0.8 °C / ± 1.6 °F
-20 to 1300 °C / -4 to 2372 °F 0.1 °C / 0.1 °F ± 0.4 °C / ± 0.8 °F
TC-L -200 to 900 °C / -328 to 1652 °F 0.1 °C / 0.1 °F ± 0.4 °C / ± 0.8 °F DIN-43710
TC-C 0 to 1500 °C / 32 to 2732 °F 0.1 °C / 0.1 °F ± 0.5 °C / ± 1.0 °F W5Re / W26Re
1500 to 2320 °C / 2732 to 4208 °F 0.1 °C / 0.1 °F ± 0.7 °C / ± 1.4 °F
The values of accuracy cover one year period and for a temperature range between 20 and 26 °C. Outside this range, the thermal stability is 0.001% FS / °C with reference
to 23 °C. Thermocouple with internal cold junction compensation, one must consider the error of this cold junction compensation of up to ± 0.2 °C or ± 0.4 °F.
Specifications – Pressure
(0) 0 – 250 mmH20 0,001 ± 0,05 % FS Gage pressure.
Used with air or inert gases.
(1) 0 – 1 psi 0,0001 ± 0,05 % FS
(2) 0 – 5 psi 0,0001 ± 0,025 % FS
(3) 0 – 15 psi 0,0001 ± 0,025 % FS Gage or absolute pressure.
Used with fluids (gases or liquids) compatible with 316 L stainless steel.
(4) 0 – 30 psi 0,0001 ± 0,025 % FS
(5) 0 – 100 psi 0,001 ± 0,025 % FS
(6) 0 – 250 psi 0,001 ± 0,025 % FS
(7) 0 – 500 psi 0,01 ± 0,025 % FS
(8) 0 – 1000 psi 0,01 ± 0,025 % FS
(9) 0 – 3000 psi 0,01 ± 0,025 % FS
(10) 0 – 5000 psi 0,1 ± 0,025 % FS
(11) 0 – 10000 psi 0,1 ± 0,05 % FS
(12) Others on request