The MEATEST M142 Multifunction calibrator is ideal for use in calibration laboratories and in production lines where voltage, current, resistance, capacity and frequency meters are manufactured. Load capacity of the voltage output is 30 mA – enough for most high-consumption analogue power-meters. Installed harmonic and non-harmonic shape signals allow for testing meter sensitivity to distorted signals by a signal with various crest factor. Frequency modes, suitable for calibration of multimeters and time bases of oscilloscopes, have adjustable 6-digit frequency, amplitude and duty ratio of the output signal. The calibrator can measure temperature with TC and RTD temperature sensors to show it on display or use for cold junction compensation.
Built-in multimeter
The MEATEST M142 has a built-in process multimeter that can be used independently or simultaneously with source functions which makes testing transducers, regulators and evaluation units really easy. Using a single instrument you can evaluate output signals of various types of transducers and external sensors (strain gauge, pressure, torsion, strength, etc.), read them directly from calibrator display and use them in your calibrations.
User comfort
M142 calibrator is focused on user friendliness making calibrations simple and straightforward. Each function has its dedicated chunky button and all relevant details including total relative accuracy are shown on large LCD. Range of adapters for easier terminal connection and automation possibilities, relative deviation feature that can help you with UUT error calculation and easy keyboard-controlled adjustment procedure are just a few examples to prove the point.
GPIB and RS-232 interfaces are present in both versions so the calibrator can be operated via remote control or used in WinQbase/Caliber software calibration systems including the CamOCR module for automated optical UUT readouts.
MEATEST M142 Specifications:
DC voltage | 0 – 1000 V ± 10 ppm / year |
AC voltage | 0 – 1000 V, 20 Hz – 100 kHz sinusoidal & non-sinusoidal waveforms ± 180 ppm / year |
DC current | 0 – 30 A (1000 A with 140-50 Current Coil) ± 100 ppm / year |
AC current | 1 uA – 30 A, 20 Hz – 10 kHz (1000 A with 140-50 Current Coil) sinusoidal & non-sinusoidal waveforms ± 500 ppm / year |
DC Power & Energy | 0.4 mW – 4.8 kW (240 kW with 140-50 Current Coil) 1-1999 s ± 510 ppm / year |
AC Power & Energy | 0.4 mW – 4.8 kW, 40 – 400 Hz (240 kW with 140-50 Current Coil) 1-1999 s, power factor -1 to +1 ± 635 ppm / year |
Resistance | 0 Ω – 1 GΩ, resolution 4½ digits ± 100 ppm / year |
Capacitance | 700 pF – 100 uF, resolution 4½ digits ± 0.5 % / year |
Temperature | TC BEJKNRST with cold junction compensation RTD ± 0.07°C / year |
Frequency | 0.1 Hz – 20 MHz ± 50 ppm / year |
Process Multimeter
DC voltage | 0 – 20 V mVDC resolution down to 100 nV ± 100 ppm / year |
DC current | 0 – 25 mA ± 150 ppm / year |
Resistance | 0 Ω – 2.5 kΩ, resolution 4½ digits ± 200 ppm / year |
Temperature | TC BEJKNRST RTD ± 0.1°C / year |
Frequency | 1 Hz – 15 kHz ± 50 ppm / year |