Yokogawa CA700 Pressure Calibrator

The Yokogawa CA700 Pressure Calibrator is a highly accurate and highly functional pressure calibrator specifically designed for the calibration of differential pressure and pressure transmitters, which are widely used in plant processes these days and are improving in terms of accuracy and functionality.

The CA700 has achieved the highest accuracy and widest range in the products class, enabling field calibration with greater accuracy than ever before.

 

 

 

If you would like more information regarding this product, simply email us here.
  • This field is for validation purposes and should be left unchanged.
SKU: Yokogawa CA700 Categories: , Tags: , , ,

The Yokogawa CA700 Pressure Calibrator has a basic accuracy of 0.01% of rdg on pressure measurement and 0.015% of rdg on Current / Voltage of the source and measurement portions of the CA700. The CA700 is the most accurate calibrator in the portable calibrator class. It also has the highest resolution and widest range of any hand held portable in its class with resolution of 0.001 kpa (200.00 kpa range) and a resolution of 0.0001 psi (29 psi range)

Applications for the Yokogawa CA700 include:

  • Field calibration of differential pressure and pressure transmitters
  • Pressure switch test
  • Check and I/O adjustment of an electro-pneumatic converter
  • 20 mA SIMULATE (two-wire transmitter simulator)
  • Two-wire transmitter loop check
  • Input command check and adjustment of recorders and controllers

Yokogawa CA700 Features:

  • Achieves the highest accuracy in the portable class
  • Available ranges:
    • -80 to 200 kPa
    • -80 to 1000 kPa
    • -80 to 3500 kPa
  • Basic accuracy:
    • Pressure; 0.01% rdg
    • Current/voltage; 0.015% rdg
  • 0.001 kPa resolution
  • Calibration procedures are embedded
  • IP54 dustproof and waterproof
Yokogawa CA700 Calibrator Specifications:
Pressure Measurement CA700-E-01 CA700–E–02 CA700–E–03
Pressure Type Gauge
Measurement range: Pos. : 0 to 200 kPa
Neg. : -80 to 0 kPa
Pos. : 0 to 1000 kPa
Neg. : -80 to 0 kPa
Pos. : 0 to 3500 kPa
Neg. pressure: -80 to 0 kPa
Measurement display range: To 240.000 kPa To 1200.00 kPa To 4200.00 kPa
Resolution: 0.001 kPa 0.01 kPa 0.01 kPa
Meas. accuracy 6mo.:
Positive pressure:
20 to 200 kPa:
±(0.01% of rdg + 0.003 kPa)
0 to 20 kPa: ±(0.005 kPa)
(0.01% of rdg + 0.04 kPa) ±(0.01% of rdg + 0.15 kPa)
Meas. accuracy 6mo.:
Negative pressure:
±(0.2% of rdg + 0.080 kPa) ±(0.2% of rdg + 0.08 kPa) ±(0.2% of rdg + 0.08 kPa)
Meas. accuracy 1yr.:
Positive pressure:
20 to 200 kPa:
±(0.01% of rdg + 0.010 kPa)
0 to 20 kPa: ±0.012 kPa
±(0.01% of rdg + 0.08 kPa ±(0.01% of rdg + 0.30 kPa)
Meas. accuracy 1yr.:
Negative pressure:
±(0.2% of rdg + 0.090 kPa) ±(0.2% of rdg + 0.09 kPa) ±(0.2% of rdg + 0.09 kPa)
Response time: 2.5 s or less
Allowable input: 2.7 kPa abs to 500 kPa gauge 2.7 kPa abs to 3000 kPa gauge 2.7 kPa abs to 4500 kPa gauge
Internal volume: Approx. 6 cm3
Temperature coefficient: ±(0.001% of reading + 0.001% of full scale)/°C or less
Influence of positional setup: Zero-point drift ±0.3 kPa or less
Measurement fluid: Gas and liquid (non-corrosive, non-flammable, non-explosive, and non-toxic fluids)
Measurement fluid temperature: -10 to 50°C
Pressure sensor: Silicon resonant sensor
Pressure sensor element: Diaphragm
Pressure display units: kPa and other units
(Pa, hPa, MPa, mbar, bar, atm, mmHg, inHg, gf/cm2, kgf/cm2, mmH2O@4°C, mmH2O@20°C, ftH2O@4°C, ftH2O@20°C, inH2O@4°C, inH2O@20°C, Torr, psi )
Input port: Rc 1/4 or 1/4 NPT female thread (selectable
Measurement unit material: Diaphragm: Hastelloy C276 and input port: SUS316

 

Yokogawa CA700 General Specifications:
Display: Dot matrix LCD (320×240 dots) c/w LED Backlight
Display refresh rate; Approx. 300ms (3 times/sec)
Warm-up time: Approx. 5 minutes
Power Voltage: Six alkaline AA batteries
Battery life: Approx. 35 hours when measuring current with the 24 V loop power supply OFF and approx 10 hours with the 24 V loop power supply ON
Auto power-off: Approx. 60 minutes (the function can be disabled)
Insulation resistance: 100 MΩ or more (500 VDC) between the input terminal and case and between the input port and case
Withstand voltage: 500 VAC for 1 minute between the input terminal and case and between the input port and case
Protection grade: IP54 dustproof and waterproof structure
Dimensions: Approx. 264 (W) x 188 (H) x 96 (D) mm, excluding protrusions
Weight: Approx. 2kg (including batteries
Compliance standards: Safety: EN61010-1, EN61010-2-030, contamination class 2
EMC: EN61326-1 Class A, EN55011 Class A Group 1
Operating temperature: -10 to 50°C and 20 to 80%RH (no condensation)
Storage temperature: -20 to 60°C and 20 to 80%RH (no condensation)
Interfaces: Select and switch between USB A mass-storage device, USB mini-B communication device class, and mass storage class
External sensor: A dedicated external sensor can be connected via a connector. (Planned to be released in the future
Accessories: A set of 1.7m long black and red lead wires with alligator clips for generation and measurement, six alkaline AA batteries, R1 1/4” – 1/8” NPT female thread x 1, ferrite core x 2, R 1/4” – 1/4” NPT female thread x 1, accessory case, instruction manual (CD), startup guide, shoulder strap